SPECTRAL
Range: 185 to 900 nm
Accuracy: ±0.5m
Bandwidth: Auto select in 8 steps, 0.125, 0.25, 0.5, 0.75, 1.0, 1.5, 2 & 2.5nm.
Readability: 0.1nm
Repeatability: 0.2nm
PHOTOMETRIC
Range: 0 to 2 Abs
Accuracy: ± 0.010 Abs. at 1.0Abs
Light Source: Hollow cathode lamp (HCL)
Burner: Titanium construction, Premix, laminar flow, 100mm
Fuel: Acetylene (C2H2)
Oxidant: Air/Nitrous Oxide (N2O)
Flow control: Integral gas control system with panel flow meters.
Nebulizer: Pneumatic, with 90% platinum and 10% iridium alloy capillary
Sample Rate: 3 to 8 ml/min (adjustable)
Monochromator: Czerny Turner 1/3m with 1800 lines/mm holographic diffraction grating and hard coated optics
Detector: Wide range Photo Multiplier Tube (PMT)
Readout: 4 line 20Character Dot Matrix backlight LCD Module
Power Supply:
HC Lamp: 0 to 30mA, modulated
D2 Lamp: 15 to 300mA, modulated
POWER REQUIREMENT: 230V ±10%, 50Hz, 1ф, Max. 200VA